18 GMS810-/GMS811-FIDOR · Operating Instructions · 8013911 V1-0 · © SICK AG
Product Description
Subject to change without notice
2.7 Information on operating gases
2.7.1 Instrument air
The instrument air is used as:
Induction air for the ejector.
Control air for pressure control.
Combustion air for the FID (for the “internal catalyst” option).
2.7.2 Combustion air
Instrument air (for measuring range > 500 mgC/m
3
).
Internal catalyst (option) (for measuring range <= 500 mgC/m
3
).
External catalyst (for measuring range <= 500 mgC/m
3
).
2.7.3 Fuel gas
Hydrogen.
Recommendation:
Monitor the pressure of the combustion gas with a pressure monitor which emits an elec-
tric signal at a certain minimum pressure (e.g. 10 bar).
2.7.4 Test gas
Zero gas: From internal catalyst or external.
Reference gas:
Recommendation: Propane in synthetic air.
Concentration: Approx. 75% of full scale reading.
Recommendation:
Monitor the pressure of the reference gas with a pressure monitor which emits an electrical
signal at a certain minimum pressure (e.g. 10 bar).
This Section contains general information on operating gases.
Quality of operating gases
p. 74, »Gas supply«